PerkinElmer
check quantity

PC3X Peltier Cyclonic Spray Chamber Heater/Cooler for Optima 2x00/3x00/4x00/5x00/7x00 DV/8x00

This PC3X Peltier heated and cooled inlet system, which includes a quartz cyclonic spray chamber, is used with Optima 2x00 DV, 3x00 DV, 4x00 DV, 5x00 DV, 7x00 DV, and 8x00 ICP-OES series instruments.

You successfully added item(s) to your cart

Part Number
List Price
Your Price
Quantity
N0786019
4261.00 EUR
more
Buy Now

Please enter valid quantity

Please login to add favorites

NULL OR EMPTY CART

Detail Information

The PC3X is a compact, Peltier-heated or cooled inlet system which incorporates the ESI cyclonic spray chamber. In cooling mode the PC3X heats or cools the outer walls of the cyclonic spray chamber, reducing the water or solvent vapor loading on the plasma, resulting in enhanced stability and performance. In heating mode, sensitivity can be improved for low flow rate applications.

  • Temperature control from -10 °C to +80 °C
  • Thermal stabilization of spray chamber improves long-term stability
  • Last set temperature saved for stand-alone operation
  • USB or Bluetooth® connectivity
  • Remote monitoring and control

This sample introduction kit includes:

  • PC3X Peltier Heater/Cooler and Power Supply
  • Baffled Cyclonic Quartz Spray Chamber
  • Connector for ESI Injector

Specifications

Model Compatible With Optima 3x00 DV, Optima 4x00 DV, Optima 2x00 DV, Optima 5x00 DV, Optima 7x00 DV, Optima 8x00
Product Brand Name Peltier
Product Group Temperature Controlled Spray Chambers
Technology Type ICP-OES