PerkinElmer uses cookies to ensure that we give you the best experience possible on our website. This may include cookies from third party websites. If you continue without changing your settings, we will assume that you consent to receive cookies from this website. You can change your cookie settings at any time. To learn more, please review our cookie policy, which includes information on how to manage your cookies.

Mean Arterial Pressure (MAP) and Pre-eclampsia

There is evidence that in a high proportion of pregnancies predisposed to develop pre-eclampsia the maternal mean arterial pressure (MAP) is increased at 11 to 13 weeks1. Combination of MAP with other markers facilitates a more effective method of screening for pre-eclampsia.

How to Measure Mean Arterial Pressure (MAP)

The blood pressure should be measured using automated devices, validated for pre-eclampsia (see list below).

DeviceModelValidation criteria*

OMRON

MIT Elite (HEM-7300-WE)

A/A2

Microlife

3BTOA-2

A/B3 or B/B4

Microlife

WatchBP Home

B/A5

Microlife

3AC1-1

B/A6

OMRON

M7 (HEM 780)

B/B7 or B/A8

GE Carescape

(Old name Dinamap ProCare 400)

A/B9

*Validation criteria recommended by British Hypertension Society (BHS). The device has to achieve a grade A or B for both SBP and DBP to be recommended for clinical use.

Positioning


  • The woman should be in a comfortable seated position with her arms supported at the level of the heart and legs uncrossed.
  • Cuff size should be selected depending on the mid-arm circumference.
  • Neither observer nor patient should talk for 5 min before or during the measurement.

Mean arterial pressure positioning

Measurement


  • The blood pressure (BP) should be measured in both arms simultaneously.
  • Series of recordings should be made at 1-minute intervals.
    • It is recommended10 to take the measurements until variations between consecutive readings fall within 10 mm Hg systolic (SBP) and 6 mm Hg in diastolic pressure (DBP) in both arms and then the last two measurements are used10.
    • Alternatively, the minimum of 2 readings can be used11.

How is The Mean Arterial Pressure (MAP) Calculated?

MAP is calculated from the systolic and diastolic readings:

MAP systolic and diastolic readings

References

  1. Poon LC et al. (2008) Mean arterial pressure at 11(+0) to 13(+6) weeks in the prediction of pre-eclampsia. Hypertension. 51(4):1027-33. Epub 2008 Feb 7. 
  2. Pregnancy Hypertension: Volume 2, Issue 4 , Pages 411-415, October 2012. Y. Chung, M.C. Brochut, A. de Greeff, A.H. Shennan. Clinical accuracy of inflationary oscillometry in pregnancy and pre-eclampsia: Omron-MIT Elite.
  3. Reinders A et al. (2005) An International Journal of Obstetrics & Gynaecology 112 : 915-920 An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A. BJOG.
  4. Nouwen E, et al. Hypertens Pregnancy. 2012;31(1):131-9. doi: 10.3109/10641955.2010.544799. Epub 2011 Feb 18.Validation of the Omron M7 and Microlife 3BTO-A blood pressure measuring devices in pre-eclampsia.
  5. Chung Y, de Greeff A, Shennan A. Hypertens Pregnancy 2009;28:348-59.Validation and compliance of a home monitoring device in pregnancy: microlife WatchBP home.
  6. The Microlife 3AC1: An accurate blood pressure measurement device in pregnancy and pre-eclampsia. de Greeff A, Reggiori F, Anthony J, Shennan A. J Hypertens 2006;24.
  7. Blood Pressure Monitoring 2009, 14:37-40 Annnemarie de Greeff, Zara Beg, Zahra Gangji, Edwina Dorney and Andrew H. Shennan. Accurracy of inflationary versus deflationary oscillometry in pregnancy and pre-eclampsia: Omron-MIT versus Omron-M7.
  8. Nouwen E, et al. Hypertens Pregnancy. 2012;31(1):131-9. doi: 10.3109/10641955.2010.544799. Epub 2011 Feb 18.Validation of the Omron M7 and Microlife 3BTO-A blood pressure measuring devices in pre-eclampsia.
  9. Hypertens Pregnancy (2010)  29:198-205,210 Accuracy Assessment of the Dinamap ProCare 400 in Pregnancy and Pre-eclampsia. De Greeff et al.
  10. National Heart Foundation of Australia: Hypertension Management Guide for Doctors 2004.
  11. Protocol for measurement of mean arterial pressure at 11-13 weeks' gestation. Poon LC, Zymeri NA, Zamprakou A, Syngelaki A, Nicolaides KH. Fetal Diagn Ther. 2012;31(1):42-8.

 